Scale Calibration
Add Point waits 10 seconds, skips the first second of bounce, then saves the median millivolt. You only type pounds. ZERO on the dashboard still only moves empty.
Saved spanNeed 0 lb and a known weight
Cal empty / live zero—
How to calibrate
- Start empty. Enter 0, ADD POINT, wait the capture to finish — that is the empty millivolt.
- Put a known weight on, then hit ADD POINT. Do not wait for STABLE. Capture skips bounce, then takes the median.
- Check the saved millivolt next to each weight. A 25 lb point that sits far off the 10 / 35 line is a bad capture — delete and redo it.
- Day to day: dashboard ZERO on empty. That does not change the span.
Add Calibration Point
Puts the weight on, then ADD POINT. Capture is ~10 s (median mV after a 1 s skip). STABLE is not required.
Saved Calibration Points
0